Clarity INT9 A/D Betriebsanweisung Seite 15

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 60
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 14
Clarity - Getting Started Installation
15
Fig. 2. System Configuration dialog
Set the number of instruments in the Number
of Instruments field
.
Note: A larger number of instruments can be set than the
amount you have purchased. You will not be able
to measure on the surplus instruments (indicated
by a blue symbol of the curve on the tab), but you
may use them e.g. for “offline” preparation of
methods.
Set the corresponding type of chromatograph
(GC/LC/EA/GPC...) for each instrument with
the Instrument Type switch
.
From the Setup Control Modules list of
installed equipment on the left drag the
equipment according to your configuration into
the selected Clarity
Instrument x tab in the right-
hand section of the dialog using the mouse or
the
button .
Note: If necessary, add further equipment to the list on
the left using the Add button
in the lower left-
hand section of the dialog.
Seitenansicht 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 59 60

Kommentare zu diesen Handbüchern

Keine Kommentare